Affected Scanner Models
[ All intraoral scanner models (i500, i600, i700, i700 wireless, i900, i900c) ]
The following information applies to all scanner models listed above.
Symptom
Pressing the power button on the scanner does not turn off the power immediately.
Troubleshooting
1. Check Scanner Connections
Check the connections of each scanner to verify any loose contacts.
| i500 | |
1. Connection Check Reattach all cable connections. | - |
2. Check USB Cable and Power Cable Connection Ensure that the terminals are fully inserted into their respective slots. | |
| i600, i700 (Plug & Scan) / i900 / i900 classic | |
1. Connection Check Reattach all cable connections. | - |
2. Check USB-C Port Verify if the USB-C port on the laptop supports Power Delivery (PD) functionality through the laptop manufacturer's website. (Though the laptop specifications may indicate PD support, there could be compatibility differences between the i600/i700 and the laptop's USB-C port.) Try connecting to another USB-C port that supports PD. If the USB-C port’s power output is unstable, it may cause power issues. | - |
3. Change Connection Method [i600, i700] Try connecting via Standard Scan using a PD hub, USB 3.0 cable, and power adapter. | - |
| i600, i700 (Standard Scan) | |
1. Connection Check Reattach all cable connections. | - |
2. Change Connection Method Check if the laptop has a USB-C port that supports Power Delivery. If it does, try connecting using only the PD cable via Plug & Scan. | - |
3. Check Power Button Status: i500 / i600 / i700 / i700 wireless
Check if you feel a click when pressing the power button.
Ensure the button is not stuck or malfunctioning.
4. Cross-check Accessories
Try connecting using different accessories (PD cable, USB 3.0 cable, PD hub) to verify if the issue persists.