Scanner Installation
Once the software installation is complete, reboot your PC before installing the hardware.
⚠️ Caution
The package includes a power cable and a USB cable. All cables used with the scanner must be properly connected to the PC.
* Use only a USB 3.0 port when connecting the scanner to your PC.
How to connect T710/T510/T310
1. Connect the power cable.
2. Connect the USB cable via a USB 3.0 (indicated with Blue color) port. (*Important)
3. Turn on the power switch on the front of Medit 3D scanner device.
How to connect T500/T300
1. Connect the power cable.
2. Connect the USB cable via a USB 3.0 (indicated with Blue color) port. (*Important)
3. Turn on the power switch on the back of Medit 3D scanner device.
4. To verify that the scanner is installed correctly, open the Device Manager on the PC and check if it is detected.
5. To verify that the cameras are installed correctly, check if two cameras appear on the Device Manager.
Accessory preparation
1. Fill each of the two Spray Supporters with Blu-Tack.
2. Apply the Blu-Tack to the Flexible Multi-die's surface.
3. When you need to use a spray, use the spray supporter. Place the object onto the "spray supporter" and spray the scan spray. Then, place it on the "flexible multi-die."
4. Install the LED protector to cover the light coming from the scanner projector.