1. Create form information by entering tooth type, option, and notes, and then click Save & Scan.
2. Complete the scan strategy setting, and click Confirm.
- Scan model type > Impression
- Impression type > Bite T (T:tray)
- 3-axis arm > On
- Scan individual stump die > On
3. Insert the impression model to the scanner, adjust the scan depth, and start scan.
- Open the 3-axis arm connection slot and mount the 3-axis arm on the scanner.
4. Scanning the impression is complete.
- Check whether all scan images are acquired for necessary areas. If not, perform additional scan.
5. Referring to the tooth number appearing on the screen, mark the tooth area on the scan data, and click Confirm.
6. Referring to the guide image, place the prepped teeth onto the multi-die. Insert the multi-die to the scanner and start scan.
7. Scanning the multi-die has been completed.
- Check whether all scan images are acquired for necessary areas. If not, perform additional scan.
8. The impression data has been auto-aligned. Click Next.
9. Align the impression data and the prepped tooth.
- 1 point or 3 points can be used for alignment.
- Mark alignment points onto the maxilla and the occlusion respectively.
10. The alignment has been completed.
11. Remove unnecessary scan data by using editing tools, and then click Next to start the merging process.
- Tip: Select the areas to keep, invert the selection, and then remove unnecessary data.
12. Check the merged scan data, and click Exit to save and complete the case.