1. Create form information by entering tooth type, option, and notes, and then click Save & Scan.
2. Complete the scan strategy setting, and click Confirm.
- Scan model type > Impression
- Impression type > Metal & Bite T (T: tray)
- 3-axis arm > On
3. Insert the impression model to the scanner, adjust the scan depth, and start scan.
- Open the 3-axis arm connection slot and mount the 3-axis arm on the scanner.
4. Scanning the impression has been completed.
- Check whether all scan images are acquired for necessary areas. If not, perform additional scan.
- Perform the least additional scans since the bite data is used only for the occlusion.
5. Scan the maxillary impression.
- Check whether all scan images are acquired for necessary areas. If not, perform additional scan.
6. Scan the mandibular impression.
- Check whether all scan images are acquired for necessary areas. If not, perform additional scan.
7. Align the maxillary impression and the bite by manual.
-
1 point or 3 points can be used for alignment.
-
Mark alignment points onto the maxillary impression and the bite respectively.
8. Align the mandibular impression and the bite by manual.
-
1 point or 3 points can be used for alignment.
-
Mark alignment points onto the mandibular impression and the bite respectively.
9. All scan data has been aligned. Inspect the results to check that the maxilla and mandibular are accurately aligned. If necessary, you can fine tune the maxilla by using the Adjust alignment feature.
- Drag motion with the ALT key: adjusts the direction and position of the manipulator without moving the maxilla.
- Drag motion with the Ctrl key: moves the maxilla slowly.
- It is recommended to adjust the manipulator first, and then fine-tune the maxilla.
10. Remove unnecessary scan data by using editing tools, and then click Next to start the merging process.
11. Check the merged scan data, and click Exit to save and complete the case.